Coordinate measuring machine operation precautions

The operator of the coordinate measuring machine must know the knowledge of the three coordinates and have relevant experience before operating the equipment. When operating the coordinate measuring machine, the equipment should be inspected, operated and maintained according to the safety operation specifications provided by the manufacturer. The following are some of the three coordinates. Measuring machine operation precautions are for reference only. The operator of the coordinate measuring machine must know the knowledge of the three coordinates and have relevant experience before operating the equipment. When operating the coordinate measuring machine, the equipment should be inspected, operated and maintained according to the safety operation specifications provided by the manufacturer. The following are some of the three coordinates. Measuring machine operation precautions are for reference only.
(1) Try to run the machine after you are sure that you have thoroughly understood how to shut down in an emergency.
(2) The surface of the granite is used as the measurement area, and the track cannot be used as the measurement area. The machine can only be operated (the track cannot be scratched or scratched);
(3) The compressed air should not be used to clean the machine. The compressed air that has not been properly treated will cause dirt and affect the normal operation of the air bearing. A vacuum cleaner can be used to clean the machine.
(4) Keep the work surface clean and the surface of the workpiece to be tested clean.
(5) When measuring the workpiece, if the rest is in the middle, move the Z axis to the top of the workpiece to be tested (safety plane), and leave a clearance, then press the emergency stop button on the control box.
(6) Do not attempt to make the machine turn or reverse rapidly.
(7) Manually control the machine to use a lower speed and maintain a uniform speed. Do not tighten the joystick until the automatic rollback is complete.
(8) Before measuring the small hole or slot, please confirm that the retraction distance is set properly.
(9) Before running a measurement program, please check whether the current coordinate system is consistent with the coordinate system required by the program. Http://news.chinawj.com.cn Submission: