How to deal with the infusion pump alarm system

Infusion pump alarm system has six symbols shining, so that timely manner to make the appropriate treatment. Air blinks, there is air in the infusion tube, and the infusion tube is removed and re-vented during treatment. ②door shining, showing infusion pump door open. Shut off pump door and lock when handling. ③ dcc shiny, showing pipeline obstruction, check whether the pipeline folded, rolling clip closed, needle blocked. ④ lowbatt shiny, showing the infusion pump battery low, immediately connected to AC. ⑤empty shiny, showing the liquid bottle or bag empty, need to change the liquid or stop infusion. ⑥rip flow rate is flashing, the number of sensor failure? Sensor is installed incorrectly? Drop the number of cavity damage? Sensor surface contamination? Sunshine or glare. drip flow rate 2, 3 flashing, infusion set inconsistent with the actual, reset, set the actual use of the infusion set. drip4,5 flashes, wrong infusion set or infusion set incorrectly. drip flow 6 shining, showing infusion leak, with the wrong infusion device, when the infusion pump to stop working, the number of drops detected more than 10 drops trickle.